2. Sample Exchange

bar

(For a more detailed description, see the manual)

  • 1. Remove the transmitted light arm from the chamber! (LENS button light off). Severe damage will occur if this is not done before inserting samples.
  • 2. Select Sample Change from the Stage Monitor menu, and click OK.
  • 3. Blank the electron beam. (PCD button lit)
  • 4. Seal the airlock lid and press the green button to evacuate the airlock.
  • 5. When evacuation is complete (when the green light goes off wait 4-5 minutes), slowly open the isolation valve between the column and the airlock, keeping the vacuum less than 5x10-5 torr.
  • 6. Retrieve the sample holder using the sample exchange rod.
  • 7. Close the isolation valve and press the green button to vent the airlock.
  • 8. Open the airlock by lifting and pulling on the lid, remove and replace samples. Remember to wear gloves, and to be sure you have oriented the sample holder properly!
  • 9. Seal the airlock lid, and press the green button to evacuate the airlock.
  • 10. When the airlock is evacuated (when the green light has gone out wait 4-5 minutes), slowly open the isolation valve, keeping vacuum less than 5x10-5 torr.
  • 11. Replace the sample holder slowly, so as not to lose y-axis calibration.
  • 12. Close the isolation valve and press the green button.
  • 13. Click on the OK button in the Sample Exchange confirmation window.
  • 14. Make sure the green button is lit before proceeding.